Examples

Coaxial Sample Holder Measurement Repeatability (0.5 - 18 GHz, PTFE)

Measurement Results

  • Calibration was performed before each measurement, and the PTFE sample was measured five times.

The graph above shows the measurement results for the relative permittivity and dielectric loss tangent of PTFE. The results show excellent repeatability, with variation below 2%. This confirms that the coaxial sample holder measurement system provides stable measurements and is suitable for quality control and comparative evaluation.

Key Point!

  • Excellent repeatability, with variation below 2%!
FEATURED PRODUCTS
CONTACT

For inquiries about our measurement services, please use this contact form.

System Configuration

  • 7 mm Coaxial Sample Holder CSF-APC7 (EM labs)
  • Measurement Software (EM labs)
  • Network Analyzer: P5007A (Keysight Technologies)
  • Calibration Kit 7 mm: 85050D (Keysight Technologies)

Samples

  • PTFE (5mm)

-Examples